Types
of spectrographs and their functioning |
Each spectrograph splits electromagnetic radiation (within the wavelenght workspace and the available resolution) in quasi-monochromatic radiation. We work in the visible range, ie at wavelengths between roughly 4000 and 7000 Angstroms = 400 to 700 nm (0.0000004 to 0.0000007 meters). The "white" light consists of a stream of photons (= smallest units of electromagnetic radiation = small localized wave packets, each representing a given energy unit). The photonic energy determines the color = wavelength of the photon. The most important part of a spectrograph is the dispersing element. It splits the photons of different color. For visible light spectroscopy prisms and optical gratings are commonly used. A prism is exploiting the difference in refractive power of glass for photons of different energy. For details, see http://en.wikipedia.org/wiki/Dispersive_prism. Optical gratings work by light diffraction and interference caused of regularly repeating patterns (think of shimmering color CD's). For details, see http://en.wikipedia.org/wiki/Diffraction_grating. For various types of spectrographs see here. Some optical basics. |
|
![]() |
The simplest spectrograph consists of a dispersive element (ie, prism
or grating) and a camera (= imaging optics) and a detector. The detector
can be your eye, then this is called a spectroscope. For faint objects
you will use more sensitive detectors, especially before the photographic
plate, now prefer electronic detectors such as CCD or digital cameras.
See also http://www.astrosurf.com/~buil/staranalyser/obs.htm
|
![]() |
In order to be able to Faint Object Spectroscopy use a light-harvesting
machine. This is the telescope.
It concentrates the light falling through its aperture in a small image
of the star in the focus. Subsequently, the light of the small (usually
15 to 60 um) star disk in the focus is again converted into parallel light
by the collimator. However, this parallel beam is much narrower as the
opening of the telescope and therefore more concentrated = brighter (see
graph). Thus, while the Staranalyser directly works with the parallel incoming starlight (works without a "concentration") the combination of telescope and collimator serves only the light collection and concentration. Strictly speaking, telescope and collimator are one instrument (although the collimator is placed in the spectrograph). In front parallel starlight falls within the telescope aperture and behind the collimator, it occurs in parallel and compacted again. Here one could also install a simple unit such as the Staranalyser (transmission grid + camera + detector) and would thus produce a simple spectrograph available, which is able already to show the spectrum of a very faint star at low resolution.
|
| Type of dispersion element |
Advantages |
Disadvantages |
| Prisms |
Virtually no light loss (High Efficiency) |
High weight of each prism High resolution can only be achieved by cascading many prism --> high weight Nonlinear dispersion function |
| Gitter |
Lightweight and small |
Higher light losses due to distribution
of light intensity over several orders of diffraction Clean the surface is very sensitive |
|
In the past 10 years, the reflection grating enforced in the amateur field. Meanwhile even enter the high-resolution Echellespektrographs into the amateur scene. They have the great advantage that the entire wavelength spectrum is mapped 350 to 700 nm at a time, divided into about 30 to 50 "orders" = part of spectra, which can then be concatenated by the software to a long range spectrum. For examples, see the Baches and the eShel of shelyak. However, it should be stressed here again that the measurement task defines which type of spectrograph the "right one". |
|
![]() |
The well-suited for self-construction and in many ways for the amateur
valuable type of spectrograph is the classical grating spectrograph
with a rotating reflection grating (see opposite). The camera is due
to the high resolution (wide "fan"), only a portion of the optical wavelength
range from (in the graphic the green light). By rotating the grating the
desired spectral range can be set on the detector. The angle between the two optical axes of the collimator and the camera (= total diffraction angle) plays an important role. It should be small, so that grids can be used with a high stroke count (high dispersion). A 90-degree design, as it is realized in the Dados spectrograph, yields a low resolution spectrograph (up to 900 lines / mm grating at maximum). Gratings with a higher dispersion are no longer usable. My first spectrograph had a total refraction angle of 45 °, which also had not been optimal. I was able to use a grating with 1200 lines / mm (R = 6000). |